3 • Operation and Maintenance
Doc. 065051-03 1/08
69
also lists the maximum injection volume that can be specified in
the Chromeleon or Chromeleon Xpress sequence or in the front panel
schedule.
Notes
•
To minimize the backpressure applied to the sample syringe pump,
use the largest ID tubing possible to make the injection loop.
•
If a 1.5 mL vial is used, the maximum loop size is 500
μ
L and the
maximum injection volume is 1100
μ
L, regardless of the syringe size.
•
For syringes that are 1,000
μ
L or smaller, the maximum loop size and
injection volume requires the syringe to aspirate from the sample vial
multiple times. This can result in reduced precision. For maximum
precision, use a loop size and injection volume that require only a
single sample aspiration.
Sample
Syringe
Sampling
Needle Assembly
Maximum Sample
Loop Size
Maximum
Injection Volume
≤
1,000
μ
L
1.2 mL
500
μ
L
a
a.
Requires a 1.5 mL vial or a 2 mL deep well plate.
1,100
μ
L
5 mL 8.5 mL
4,000
μ
L
b
b. Requires a 10 mL vial.
4,700
μ
L
10 mL
8.5 mL
7,000
μ
L
b
8,000
μ
L
Table 3-9. Maximum Sample Loop Sizes and Injection Volumes
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...