AS Autosampler Operator’s Manual
314
Doc. 065051-03 1/08
18. Select the following options:
DISPENSE
:
1
μ
L
SYRINGE SPEED
:
1
ACTION
:
DISPENSE
19. Press
Enter
.
The syringe dispenses 1
μ
L.
20. Continue pressing
Enter
, counting each key press,
until a tiny drop of liquid
appears at the end of the needle seal line
. After each key press, wait until
the sound of the syringe motor stops before pressing
Enter
again.
21. Record the volume entered (including the volume in
).
Also, notice the size of the liquid drop. This will be compared to the size of
the liquid drop in
22. Select the following options:
ASPIRATE
:
60
μ
L for normal mode;
120
μ
L for sequential mode (to empty the
needle seal line and diverter tubing)
from
NEEDLE
SYRINGE SPEED
:
3
ACTION
:
ASPIRATE
23. Press
Enter
.
24. Select
VIAL#
:
FLU
and press
Enter
.
The needle arm moves to the flush port.
25. Select the following options:
SYRINGE SPEED: 5
ACTION:
EMPTY
26. Press
Enter
.
27. Select the following options:
from
FLUSH
SYRINGE SPEED
:
5
ACTION
:
FILL
28. Press
Enter
.
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...