Example Program for Normal Mode
The following example Chromeleon or Chromeleon Xpress program
shows the AS commands for sample loading and injecting when the
autosampler is in normal mode (full-loop, partial-loop, or partial-loop,
limited sample).
3.6.6
Normal, Large-Loop Injections
The injection sequence for a normal, large-loop injection is the same as
described for a normal, full-loop injection (see
when a large loop (200.0 to 7000.0
μ
L)
is installed, the AS draws a set
amount (500
μ
L) of extra sample from the vial for flushing the loop,
instead of drawing a factor of the loop volume (see
for
details).
0.000
Load
;Switch the inject valve to Load
and load sample onto the loop
Wait
CycleTimeState
;Wait for the end of the specified
Cycle Time
Inject
;Switch the inject valve to Inject
and sweep the sample onto the
column
Wait
InjectState
;Wait for the injection to finish