74
Doc. 065051-03 1/08
Setting Up a Partial-Loop Injection
1. On the
SYSTEM PARAMETERS
screen (see
), select
NORMAL
in the
SAMPLE MODE
field.
2. In the Chromeleon or Chromeleon Xpress sequence (see
or in the front panel schedule (see
), enter an injection
volume that is less than the volume of the sample loop installed on the
injection valve.
3. In the Chromeleon or Chromeleon Xpress program (see
or on the
SYSTEM PARAMETERS
screen, enter a cut volume greater
than zero.
NOTE For the best accuracy when performing partial-loop
injections, specify an injection volume equal to half
the loop size or less.
Figure 3-21. Program Wizard: Specifying the Cut Volume