4 • Troubleshooting
Doc. 065051-03 1/08
115
•
[415] Cycle time expired before INIT step was completed.
When the cycle time expires, the AS ignores the cycle time and
delays injection until the
INIT
step is complete. See
for details.
Possible Causes:
The selected cycle time was too short.
Actions:
Select a longer cycle time (see
).
•
[418] DSP not able to home XYZ motor arm
Possible Causes:
•
Physical blockage of the sampling arm
•
DSP malfunction
Actions:
1. Turn off the AS power.
2. Check for an obstruction in the autosampler compartment.
3. Check for freedom of movement in all three axes of the sampling
arm.
4. Turn on the power again. If the problem persists, contact Dionex.
•
[421] Timed events step took so long to execute that at least one
subsequent step was missed.
Possible Causes:
•
Insufficient time allowed for loading the sample loop
Actions:
•
On the method’s
TIMED EVENTS
screen, set the injection valve to
the load position in the
INIT
step (the default).
•
If the valve is not set to load in the
INIT
step, be sure to allow
enough time between the
LOAD
and the
INJECT
parameters
for the
loop to be completely loaded. The time needed depends on the
sample syringe volume and the selected syringe speed (see
).
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...