AS Autosampler Operator’s Manual
176
Doc. 065051-03 1/08
A.1.3
Valves (Optional)
A.1.4
Injection
Injection
One or two, two-position, six-port, electrically-activated Rheodyne
valves, PEEK
Diverter
One two-position, six-port, electrically-activated Rheodyne valve,
PEEK
Injections per Vial
1 to 99
Vial Sizes Available
0.3 mL polymer vials
1.5 mL glass vials
10 mL PolyVials
Well Plate Sizes
Available
Deep-well plate, 96 wells, 2.0 mL
Micro-well plate, 96 conical (V-shaped) wells, 0.45 mL
Micro-well plate, 96 round (U-shaped) wells, 0.5 mL
Variable Volume
Range
If the sample syringe is 1,000
μ
L or less:
1.0 to 99.9
μ
L in 0.1
μ
L increments; 100 to 1,000
μ
L in
1
μ
L increments (normal or sequential mode)
100 to 1,000
μ
L in 1
μ
L increments (concentrate or sequential
concentrate mode)
If the sample syringe is 5,000
μ
L:
1.0 to 99.9
μ
L in 0.1
μ
L increments; 100 to 4,700
μ
L in
1
μ
L increments (normal or sequential mode)
100 to 4,700
μ
L in 1
μ
L increments (concentrate or sequential
concentrate mode)
1,000 to 4,700
μ
L in 1
μ
L increments (simultaneous mode)
If the sample syringe is 10,000
μ
L:
1.0 to 99.9
μ
L in 0.1
μ
L increments; 100 to 8,000
μ
L in 1
μ
L
increments (normal or sequential mode)
100 to 8,000
μ
L in 1
μ
L increments (concentrate or sequential
concentrate mode)
1,000 to 8,000
μ
L in 1
μ
L increments (simultaneous mode)
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...