AS Autosampler Operator’s Manual
304
Doc. 065051-03 1/08
D.7 Selecting Syringe Speeds for Front Panel Operation
For front panel operation, syringe speeds are set on the
SYSTEM PARAMETERS
screen (see
NOTE The syringe speed settings described in this section are
used only if the AS is controlled exclusively from the
front panel. When the autosampler is controlled by
Chromeleon or Chromeleon Xpress (even if it is not
currently connected to the software), the AS uses the
syringe speeds selected in the software (see
Two syringe speed parameters are available:
METHOD
SYRINGE SPEED
and
FLUSH SYRINGE SPEED
.
•
The
METHOD
SYRINGE SPEED
is used for the following operations:
•
Aspirating and dispensing sample during loop or concentrator loading
•
Aspirating and dispensing sample during sample prep operations, except
for mixing and dilution
•
Aspirating from and dispensing to the flush reservoir during sample prep
operations and loop or concentrator loading
•
Aspirating an air bubble
•
The
FLUSH SYRINGE SPEED
is used for the following operations:
•
Flushing
•
Reagent priming after concentrator loading
•
Reagent priming and flushing after concentrator loading (when a
reservoir is the specified source).
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...