3.6.4
Guidelines for Sample Volumes Used, Loop Sizes, and
Injection Volumes
Sample Volumes Used for Each Type of Sample Loop Injection
shows the volume of sample the AS aspirates from the sample
vial or well plate well and the volume of sample injected onto the column.
See
for descriptions of sample loop injection types and the
criteria used to determine which type of injection is performed.
The following tables give several examples for full-loop, large-loop, and
partial-loop injections. The examples show the amount of sample used
and injected for various loop sizes and injection volumes.
Injection Type
Sample Volume Aspirated
from Vial or Well
Sample Volume
Injected
Full-loop
(2.5–16.9
μ
L loop)
Injection volume or
2.5
×
loop size + 25
μ
L
(whichever is greater)
Loop size
Full-loop
(17.0–199.9
μ
L loop)
Injection volume or
4
×
loop size
(whichever is greater)
Loop size
Large-loop
(200.0–7000.0
μ
L loop)
Injection volume or
1
×
loop size + 500
μ
L
(whichever is greater)
Loop size
Partial-loop
Injection volume
+
2
×
cut
volume
Injection volume
Partial-loop, limited-
sample
Injection volume
Injection volume
Table 3-5. Sample Volumes Used for Each Injection Type
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...