AS Autosampler Operator’s Manual
168
Doc. 065051-03 1/08
about 1.5 cm into the port. If the drive shaft position is incorrect, you will not
be able to insert the paper clip more than about 1 cm into the port. In this case,
rotate the valve 180 degrees and recheck the position.
5. Each valve port must contain a white Teflon washer. Check the inside of each
port and insert a washer (P/N 061315), if one is not already present. The
washers are supplied with the syringe valve and with the
AS
Preventive
Maintenance Kit (P/N 060581).
6. Holding the syringe at the top fitting, screw it into the bottom of the valve at
port S and tighten fingertight.
7. With valve ports L and B at the 12 o’clock position and the syringe at the
6 o’clock position, push the valve onto the syringe drive shaft.
8. Replace the two valve thumbscrews and partially tighten.
9. Carefully align the syringe piston with the connecting screw mount on the
syringe drive.
10. Insert the syringe drive connecting screw and tighten fingertight. Finish
tightening the valve thumbscrews.
11. Reconnect the valve liquid lines. Tighten the fittings fingertight.
12. On the
MENU of SCREENS
,
press
5
and
3
to go to the
PLUMBING
CONFIGURATION
screen. Set the initial position (see
13. On the
PLUMBING CONFIGURATION
screen, move the cursor to the
PREP
SYRINGE VOLUME
field. Press
Select
Δ
or
Select
∇
to select the syringe
volume. Press
Enter
.
If a fitting leaks when operation is resumed, tighten it slightly. Do not
overtighten. Overtightening can damage the inside of the valve.
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...