B • Installation
Doc. 065051-03 1/08
211
NOTE The upper limit for the loop size depends on the
volume of the sample syringe installed: 500
μ
L
(if the syringe volume is
≤
1 mL), 4,000
μ
L (if the
syringe volume is 5 mL), and 8,000
μ
L (if the
syringe volume is 10 mL).
5. For the simultaneous or sequential mode, move the cursor to the
LOOP SIZE V2
field and enter the size of the sample loop installed on
the second injection valve.
Select the Wait Operation (for Front Panel Operation Only)
NOTE Do not complete this section if the AS will be
controlled by Chromeleon or Chromeleon
Xpress. The software enables the wait function
automatically and ignores the setting on the
SYSTEM PARAMETERS
screen.
1. Go to the
MODULE SETUP MENU
and press
5
to go the
SYSTEM
PARAMETERS
screen.
2. Move the cursor to the
WAIT
field and press
Select
Δ
or
Select
∇
to
select either
ENABLED
or
DISABLED
.
•
If the AS will be controlled by another instrument through TTL
or relay connections, set the
WAIT
field to
ENABLED
.
•
If the AS will control other modules (system master) through
TTL or relay connections, set the
WAIT
field to
DISABLED
3. Press
Enter
.
Figure B-45. System Parameters Screen
Help Prompt
SYSTEM PARAMETERS
WAIT:
ENABLED
METHOD SYRINGE SPEED:
CUT SEGMENT VOLUME:
INJECT PORT VOLUME:
uL
uL
10
S1 43
3
SAMPLE MODE:
NORMAL
FLUSH SYRINGE SPEED:
4
uL
S2
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...