AS Autosampler Operator’s Manual
10
Doc. 065051-03 1/08
2.2.1
Keypad Button Functions
Button
Function
Stops a schedule if the AS is in Local mode, Schedule control (see
). A confirmation message appears; press
Stop
again to
abort the schedule or
Hold/Run
to resume.
If the AS is in Direct Control (see
Stop
stops
the current syringe and/or sampling arm movement.
Note:
Pressing
Stop
while the AS is in Direct Control, Locked
Remote mode (software control) (see
) is not
recommended. Use the controls on the software Control panel (see
) to stop a sequence.
Sends the needle arm to its home position. This button functions only
when the AS is in Local mode. If a schedule is in progress, it must be
on hold. If you press
Home
when a schedule is on hold, and then
resume the schedule, the needle arm returns to the location it was in
when you pressed
Home
.
Inserts a new step into a method or schedule. Move the cursor to the
leftmost field on the
SAMPLE PREP
,
TIMED EVENTS
, or
SCHEDULE
screen and press
Insert
. The new step is added above the cursor
position. See
for details about front panel operation.
Delete
performs several functions:
•
Cancels an entry that is in progress and restores the previous value.
•
Returns a field to its default value (if an entry is not in progress).
•
Deletes a line in a method or schedule. To do this, position the
cursor in the leftmost field of the line that you want to delete and
press
Delete
twice.
•
Deletes an entire method or schedule. To do this, position the
cursor in the method or schedule edit field, or in the time field of
the
INIT
step, and press
Delete
twice.
At power-up, the AS is on hold (the left LED is illuminated). After you
select a schedule to run, pressing
Hold/Run
starts the schedule (the
right LED is illuminated). When the schedule is complete, the AS
returns to hold. When running, press
Hold/Run
to pause the schedule.
This button functions only when the AS is in Local mode.
Note:
If the wait operation is enabled (see
pauses when it reaches the wait step in the method. Press
Hold/Run
to
continue. You can disable the wait (see
Stop
Home
Insert
Delete
(Hold)
(Run)
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...