4 • Troubleshooting
Doc. 065051-03 1/08
117
•
[438] Timed out waiting for home position to be found.
Possible Causes:
•
Autosampler compartment door open at power-up
•
DSP malfunction
•
Sampling arm malfunction
Actions:
•
Close the autosampler compartment door and then press the
Home
button.
•
If closing the door does not eliminate the problem, try turning the
power off and then on again. If the problem persists, contact
Dionex.
•
[439] Method/Schedule database has been corrupted. Reinitializing
database.
Possible Causes:
•
Dead battery on CPU board
•
Malfunctioning power supply
Actions:
Contact Dionex.