D • Front Panel Control
Doc. 065051-03 1/08
315
29. Select the following options:
DISPENSE
:
50
μ
L
SYRINGE SPEED
:
5
ACTION
:
DISPENSE
30. Press
Enter
.
31. Select
VIAL#
:
INJ
and press
Enter
.
The needle arm moves to the inject port.
32. Select the following options:
DISPENSE:
volume (microliters) recorded in
SYRINGE SPEED
:
3
ACTION
:
DISPENSE
33. Press
Enter
.
34. Verify that a drop of liquid appears at the end of the needle seal tubing. If no
drop appears or if the drop is a different size than previously observed, repeat
through
.
For sequential mode, switch the diverter valve to the other position and repeat
35. Select
VIAL#
:
FLU
and press
Enter
.
36. Select the following options:
SYRINGE SPEED
:
5
ACTION
:
EMPTY
37. Press
Enter
.
38. Press
Home
to move the needle arm to the home position.
39. Press
Menu
twice to go to the
MENU of SCREENS
.