AS Autosampler Operator’s Manual
234
Doc. 065051-03 1/08
TRAY
(type)
The tray type in use.
NONE, 2 mL,
P10 mL
(10 mL
plastic),
WP.5mL
,
WP 2mL
(0.5 mL
or 2 mL well plate)
LOOP
The size of the loop in microliters.
Note:
If the
AS is equipped with two injection valves, two
loop sizes are shown. If the AS is in concentrate
mode,
CONCENTRATOR
is displayed.
(Injection
Type)
The type of injection performed. This field is
updated when the initial conditions of the
method’s timed events are executed.
TTL1
TTL2
RLY1
RLY2
Provides TTL and relay control of other devices.
In Direct Control, select 1 (on) or 0 (off). In
Schedule control, set the TTL and relay fields
from the
TIMED EVENTS
screen for the method
being run.
control.
1
(on)
0
(off)
INJ VLV
Displays, and lets you set, the position of the
injection valve.
L
(load)
I
(inject)
INJ VLV 1
INJ VLV 2
(simultaneous
injection
mode only)
If the AS is equipped with two injection valves
(for simultaneous injections), these fields display,
and let you set, the position of each valve.
Injection valve #1 is the bottom valve and
injection valve #2 is the top valve. See
for details about the simultaneous
injection option.
L
(load)
I
(inject)
Diverter
Displayed when the AS is equipped with a
diverter valve for sequential injections and
sequential or sequential concentrate is the
selected sampling mode.
A
(system 1)
B
(system 2)
Screen Field
Description
Values
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...