C • Display Screens
Doc. 065051-03 1/08
253
When wait is disabled, a run command is required
to start a schedule. Thereafter, the AS performs the
schedule lines without requiring additional run
commands. Disabling wait allows the AS to act as
a system master. The AS can send TTL signals to
other modules in the system, thereby controlling
system functions (starting the pump, starting the
detector method, etc.).
NOTE This parameter is used for front panel
operation only. If the AS is controlled
by Chromeleon or Chromeleon Xpress,
the software automatically enables the
wait function and the setting on the
SYSTEM PARAMETERS
screen is
ignored.
METHOD
SYRINGE SPEED
(This parameter is displayed only when the AS is
controlled exclusively from the front panel.)
The syringe speed to use during a front panel
method run. For water, accept the default value of
3; more viscous liquids require slower speeds. For
concentrate mode, select a speed of 2.
Be sure to select a speed that does not produce
backpressure of more than 690 kPa (100 psi). See
for details. In general if you are using a
5 or 10 mL syringe, select a speed of 1 or 2.
1 (slowest) to
5 (fastest)
(default=4)
For
concentrate
mode: select 2
PIPET SYRINGE
SPEED
(This parameter is displayed only when the AS is
controlled by Chromeleon or Chromeleon Xpress.)
The syringe speed used for various pipet actions.
Accept the default value.
NOTE For Chromeleon or Chromeleon
Xpress operation, specify syringe
speeds in the program or on the
Control panel (see
1 (slowest) to
5 (fastest)
(default=4)
CUT SEGMENT
VOLUME
For partial-loop injections, specifies a volume of
sample to be “cut” (or omitted) from each side of
the sample during the injection process. See
for details about injection types.
0 to 30
μ
L
(default=10)
SAMPLE MODE
The sample delivery mode.
See the table
below.
Screen Field
Description
Values
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...