3 • Operation and Maintenance
Doc. 065051-03 1/08
67
Full-Loop Sample Volume Examples
For full-loop injections, if the injection volume is more than 4 times the
loop size, the full injection volume is aspirated. The excess aspirated
sample is flushed through the loop, but the amount injected is equal to the
loop size.
Large-Loop Sample Volume Examples
For large-loop injections, if the injection volume exceeds the loop size +
500
μ
l, the full injection volume is aspirated. The excess sample aspirated
is flushed through the loop, but the amount actually injected is equal to
the loop size.
Installed Loop
Size (
μ
L)
Injection
Volume from
Chromeleon
Sequence (
μ
L)
Sample
Aspirated from
Vial or Well
(
μ
L)
a
a. Excess aspirated sample is used to flush the loop before and after injection.
Sample Volume
Injected onto
Column (
μ
L)
10
10
40
10
10
35
40
10
10
60
60
10
25
25
100
25
25
75
100
25
25
150
150
25
Table 3-6. Full-Loop Sample Volume Examples
Installed Loop
Size (
μ
L)
Injection
Volume from
Chromeleon
Sequence (
μ
L)
Sample
Aspirated from
Vial or Well
(
μ
L)
a
a. Excess aspirated sample is used to flush the loop before and after injection.
Sample Volume
Injected onto
Column (
μ
L)
500
500
1,000
500
500
700
1,000
500
500
1,100
1,100
500
Table 3-7. Large-Loop Sample Volume Examples
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...