D • Front Panel Control
Doc. 065051-03 1/08
293
Setting the Tray Temperature
The
TRAY TEMPERATURE
parameter is available only when the sample
temperature control option is installed. Otherwise, this field displays “--”
and is not illuminated.
1. On the
METHOD SETUP
screen, position the cursor in the
TRAY
TEMPERATURE
field.
2. If the field displays “--”, tray temperature control is currently turned
off. Press
Select
Δ
or
Select
∇
to turn on tray temperature control.
3. Press numeric buttons to enter the desired temperature setting.
4. Press
Enter
.
Waiting for the Temperature to Stabilize
The
WAIT FOR TEMP STABLE
parameter delays the start of the method
until the temperature of the sample tray has reached the specified set
point(s).
1. Position the cursor in the
WAIT FOR TEMP STABLE
field.
2. Press
Select
Δ
or
Select
∇
to select either
YES
(wait) or
NO
(do not
wait).
3. Press
Enter
.