AS Autosampler Operator’s Manual
140
Doc. 065051-03 1/08
6. The sampling needle line is coiled and held by a bracket and clamp on
the upper-right side of the autosampler compartment (see
). To remove the tubing, use a flathead screwdriver to
gently pry the center bracket apart. Then, lift the tubing out of the
bracket and clamp.
7. Measure the distance from the second yellow sleeve (see
to the needle. When installing the new sampling needle assembly,
position the sleeve on the new line the same distance from the needle.
Figure 5-11. Sampling Needle Line Bracket and Clamp (Right Side of Compartment)
8. If you are replacing an 8.5 mL assembly (P/N 061267), go on to
. If you are replacing a 1.2 mL sampling needle assembly
(P/N 054271), follow the steps below to disconnect the sampling
needle line:
a. Disconnect the sampling needle assembly from port C on the
sample syringe valve.
b. Pull the ferrule fitting off the end of the tubing and remove the
bolt.
Bracket
Clamp
To Port C on Sample
Syringe Valve
Coupler
Sampling
Needle
Line
Second Yellow Sleeve
Note: This coupler is
included only on the 8.2 mL
sampling needle assembly.
Sampling
Needle
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...