B • Installation
Doc. 065051-03 1/08
201
5. Route the needle
seal line out the AS
compartment
through the slot on
the left side (see
Verify that the
tubing is not
pinched when the
door is closed.
Figure B-37. Second AS Installation:
Needle Seal Line Routed Out the Left Side of the AS
6. Align the sampling needle in the inject port (see
7. Calibrate the inject port volume (see
Figure B-36. Second AS Installation: Tubing Chase Installation
Attach tubing
chase here
Tubing chase attached
and needle seal line
routed to the DC