AS Autosampler Operator’s Manual
34
Doc. 065051-03 1/08
2.7.2
Overlapping Sample Preparation
When the wait operation is at the end of the sample prep phase (the
default), the
AS
overlaps sample preparation. This means that while data
collection is occurring for one injection, the AS performs the sample prep
steps for the next injection (see
). For a detailed description of
the schedule run flow of operations, see
Figure 2-15. Sample Prep Overlap: Wait Operation at
End of Sample Prep Phase (Default)
Sample Overlap with Chromeleon or Chromeleon Xpress
If Chromeleon or Chromeleon Xpress
is controlling the AS, front panel
settings for the wait operation are ignored. Instead, the
Sample Overlap
option determines whether sample preparation is overlapped. Sample
overlap, which is selected in the Chromeleon Server Configuration
program, is enabled by default. For instructions on how to configure the
AS in Chromeleon, see
0.0
0.0
Injection 1
Start First Method
Wait for Continue
Command
SAMPLE PREP
WASH/
FLUSH
WASH/
FLUSH
WASH/
FLUSH
NOTE:
The WAIT is ignored for
the first injection in a
Schedule.
SETUP
TIMED EVENTS
INIT
Injection 2
CONTINUE
Start Next Method
SAMPLE PREP
SETUP
TIMED EVENTS
INIT
Start Next Method
SAMPLE PREP
WAIT
WAIT
Overlap
Overlap
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...