AS Autosampler Operator’s Manual
192
Doc. 065051-03 1/08
3. Connect the needle
seal line from the AS
to port S (5) on the
injection valve (see
4. Locate the green waste
line connected to port
W (6) on the injection
valve (see
Figure B-24. Injection Valve
5. If you are connecting the AS to an ICS-3000 Detector/
Chromatography module (DC), route the waste line from the injection
valve through the side slot on the DC and then through the side slot
on the AS.
6. If you are connecting the AS to an ICS-1000/1500/2000, this green
waste line (labeled
AS TRAY
) is connected with a union to a Teflon
waste line (see
), which is routed through the tubing
chase to the chromatography system module rear panel.
7. Disconnect the green
AS TRAY
line from the union and route the line
to the drip tray in the autosampler compartment.
Figure B-25. Injection Valve Waste Line Union (ICS-1000/1500/2000)
AS
Needle
Seal Line
Waste
Line
AS TRAY
Waste Line
(Disconnect
from the union)
Union
Tubing
Chase
Teflon
Waste
Line