AS Autosampler Operator’s Manual
138
Doc. 065051-03 1/08
5.4.1
Removing the Old Sampling Needle Assembly
1. Press
Menu
,
8
, and
5
to go to the
LIQUID CONTROL
screen.
2. Move the cursor to the
VIAL#
field. Press
Select
∇
to select
FLU
and
press
Enter
.
The needle arm moves to the flush port.
3. Turn off the power and open the autosampler door.
Figure 5-9. 8.5 mL PEEK Sampling Needle Assembly (P/N 061267)
PEEK Needle
Yellow Sleeves
Ferrule
(P/N 052231)
Bolt
(P/N 052230)
Tubing
0.8-mm (0.032-in) ID
Tubing
1.5-mm (0.062-in) ID
Coupler
Ferrule
(P/N 064553)
Bolt
(P/N 052267)
(Shipped unattached.
Connect to the sample
syringe after installing
the needle assembly.)