AS Autosampler Operator’s Manual
322
Doc. 065051-03 1/08
E.2.1
Example TTL/Relay Connections
The following example connections can be used to connect AS relay and
TTL outputs to TTL inputs on an ICS-1000/1500/2000 system. These
connections allow the AS to control the following functions:
•
Injection valve position
•
Pump flow
•
Detector autozero
NOTE Refer to the ICS-1000/1500/2000 operator’s manual
for an example of how to set up the AS and the IC
system for
stand-alone operation
(operation without
Chromeleon or Chromeleon Xpress
software).
Figure E-5. Example TTL/Relay Connections: ICS-1000/1500/2000
AS50 Rear Panel Connections
ICS-1000/1500/2000
Rear Panel Connections
red: pos. 7,8,9
1
2
3
4
5
6
7
8
9 10 11 12
1
2
(+)
TTL OUT
1
2
TTL GND
(-)
1
2
3
4
TTL IN
(+)
1
2
RELAY
OUT
TTL 2 OUT (+)
TTL 1 OUT (+)
RLY 2 OUT
RLY 1 OUT
TTL 1 IN (+)
TTL 2 IN (+)
TTL 3 IN (+)
TTL 4 IN (+)
TTL
IN/OUT (-)
red pos.
3,5,6
Inject
1
2
3
4
5
6
7
8
9
10
11
12
black pos.
4,12,12
Autozero
Pump On/Off
black:
pos. 11,12,12
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...