AS Autosampler Operator’s Manual
158
Doc. 065051-03 1/08
•
Step 2:
After clicking
+1 uL
,
wait until the sound of the syringe motor
stops before clicking
+1 uL
again.
Continue clicking
+1 uL
until a tiny drop of liquid appears at the end of
the needle seal line. If you go too far and more than a single small drop of
liquid appears, click -
5 uL
. The AS will empty the line and then fill the
line with 5
μ
L less than was cleared. You can then begin clicking
+1 uL
again.
•
Step 3:
Press
Reset
and then
Test
. After clicking
Test
, verify that a drop
of liquid appears at the end of the needle seal line. If no drop appears, or if
the drop is a different size than previously observed, click -
5 uL
and
repeat the procedure.
•
Step 4:
After clicking
Done
, the new calibration value is displayed on the
panel. Write this value on the label inside the AS front door.
•
Finishing:
Reconnect the needle seal line to the injection valve. Tighten
the fitting fingertight and then tighten an additional one-quarter turn using
the 5/16-in wrench.