5 • Service
Doc. 065051-03 1/08
155
5.9
Calibrating the Inject Port Volume
Calibrate the inject port volume at the following times:
•
After replacing the needle seal assembly
•
If reproducibility between injections is poor
Two methods are available for calibrating the inject port volume:
•
Automatic calibration using Chromeleon or Chromeleon Xpress (Release 6.8
SP2 or later)
•
Manual calibration using the front panel screens
This section describes the calibration in Chromeleon or Chromeleon Xpress.
If Chromeleon or Chromeleon Xpress (Release 6.8 SP2 or later) is unavailable,
refer to
for instructions on how to use the front panel screens to
calibrate the inject port volume.
Note for Sequential Injections
For sequential injections, the volume to be calibrated includes not just the needle
seal line, but the path through the diverter valve and the diverter tubing to the
injection valve (see
Figure 5-26. Sequential Injections:
Inject Port Volume Flow Paths
1
S
2
P
P
P
Needle Seal Line
Diverter Valve
Inject Port
Diverter Line
W
L
P
C
L
S
Inject Valve #1
Inject Valve #2
W
L
P
C
L
S
Diverter Line
Needle
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...