AS Autosampler Operator’s Manual
242
Doc. 065051-03 1/08
C.4.2
Method Setup Screen
Use the
METHOD SETUP
screen to set the initial conditions and default
parameters for a method.
Figure C-9. Method Setup Screen
Screen Field
Description
Values
CYCLE TIME
The time between injections. When a method is
assigned a cycle time, its timed events are not
executed until the specified time has elapsed. See
for details.
1 to 240
minutes
(default=off)
Cycle time is ignored in the following situations:
•
If
WAIT FOR TEMP STABLE
is enabled in
the
METHOD SETUP
) and the tray or column
temperature has not stabilized by the end of
the cycle time.
•
If the wait step of the method’s sample prep
phase is enabled and a continue command
has not occurred by the end of the cycle
time.
SAMPLE
NEEDLE
HEIGHT
Controls the height of the sample needle above
the bottom of the vial.
0 to 46 mm
(default=2)
NOTE For 0.3 mL and 1.5 mL vials, do not set
the needle height greater than 2 mm. This
ensures the vent in the needle is
positioned inside the vial.
Help Prompt
METHOD:
METHOD SETUP
°C
46
min
mm
123
20
YES
CYCLE TIME:
SAMPLE NEEDLE HEIGHT:
SAMPLE TRAY TEMPERATURE:
WAIT FOR TEMP STABLE:
25
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...