AS Autosampler Operator’s Manual
308
Doc. 065051-03 1/08
7. Verify that the needle is centered over the inject port opening.
8. Adjust the alignment by changing the
X ADDRESS
and/or
Y ADDRESS
values.
Change the values in two- to five-step increments.
After each value change, press
Enter
. The needle arm moves up and then
adjusts to the new position.
•
To move the needle to the right, increase the
X ADDRESS
value.
•
To move the needle to the left, decrease the
X ADDRESS
value.
•
To move the needle forward, increase the
Y ADDRESS
value.
•
To move the needle back, decrease the
Y ADDRESS
value.
9. After each adjustment, move the cursor to the
Z ADDRESS
field (keep the
value at
180
)
and press
Enter
. The needle arm moves back down. Observe the
alignment and adjust the position again if needed.
10. When the adjustments are complete, press
Home
.
The needle arm moves to the home position.
11. Press
Enter
.
12. Position the cursor in the
X
or
Y ADDRESS
field and press
Enter
.
The needle arm moves to the selected position.
13. Recheck the alignment and adjust the
X
or
Y ADDRESS
values, if necessary.
14. Position the cursor in the
Z ADDRESS
field.
15. Enter the following value in the
Z ADDRESS
field:
230
.
16. Press
Enter
.
The needle moves into the inject port.
17. Press
Home
.
18. Press
Menu
three times to return to the
MENU of SCREENS
.
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...