5 • Service
Doc. 065051-03 1/08
137
SCREEN
) displays the number of cycles performed since the
assembly was last replaced.
Two versions of the PEEK sampling needle assembly are available. The version
used depends on the volume of the sample syringe installed:
•
If the sample syringe is 1 mL or less, a 1.2 mL sampling needle assembly
(P/N 054271) is installed (see
•
If the sample syringe is 5 or 10 mL, an 8.5 mL sampling needle assembly
(P/N 061267) is installed. The 8.5 mL assembly includes the same parts as the
1.2 mL assembly, plus a coupler and an additional length of tubing to increase
the volume (see
NOTE The PEEK sampling needle assemblies use a different
bolt and ferrule than previous sampling needle
assemblies. If you remove the PEEK needle from the
0.8-mm (0.032-in) ID tubing, be sure to use the correct
fitting bolt (P/N 052667) and ferrule (P/N 064553) to
reattach the needle to the tubing.
Figure 5-8. 1.2 mL PEEK Sampling Needle Assembly (P/N 054271)
PEEK Needle
Yellow
Sleeves
Ferrule
(P/N 052231)
Bolt
0
(P/N 05223 )
Ferrule
(P/N 064553)
Bolt
(P/N 052267)
Tubing
0.8-mm (0.032-in) ID
(Shipped unattached.
Connect to the sample
syringe after installing
the needle assembly.)
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...