3 • Operation and Maintenance
Doc. 065051-03 1/08
79
Setting Up a Simultaneous Injection
The AS must be equipped with the simultaneous injection option in order
to perform simultaneous injections. The following setup steps are
required:
•
Follow the instructions in
to connect the AS to two
injection valves (either in the autosampler compartment or two IC
systems). In this configuration, the inject port tubing is connected to a
Y-connector and the legs of the Y are connected to the inject valves in
both systems (see
). The tubing from the Y to each valve
must be the same length.
•
Install a sample syringe of 5,000 or 10,000
μ
L.
•
Enter the sample syringe size of 5,000 or 10,000
μ
L on the
PLUMBING CONFIGURATION
screen (see
•
On the
SYSTEM PARAMETERS
screen (see
), select
SIMULTANEOUS
in the
SAMPLE MODE
field.
•
In the Chromeleon or Chromeleon Xpress sequence (see
or in the front panel schedule (see
), enter an injection
volume that is large enough to allow flushing the inject port, flushing
the tubing from the inject port to each valve (25
μ
L per inch of
tubing), plus enough to flush and fill both injection loops (this is
about 5 times the combined loop sizes).
Example Simultaneous Injection Parameters:
45
μ
L = inject port volume
500
μ
L = 20 inches of connector tubing (combined from both
systems) x 25
μ
L
625
μ
L = 5 x (100
μ
L loop #1 + 25
μ
L loop #2)
total = 1170
μ
L
The specified injection volume is loaded into the sample syringe and
then delivered equally to the two sample loops.
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...