AS Autosampler Operator’s Manual
58
Doc. 065051-03 1/08
3.5
Specifying Injection Volumes
The range of injection volumes the AS can deliver to the injection valve depends
on the size of the sample syringe installed on the AS and on the sampling mode
selected on the
SYSTEM PARAMETERS
screen (see
lists the volume range for each syringe size and sampling mode.
You can enter the injection volume in the Chromeleon or Chromeleon Xpress
sequence (see
) or in the front panel schedule (see
). For volumes less than 100
μ
L, specify the value in 0.1
μ
L
Syringe Size(s)
Sampling Mode
Injection Volume
100–1,000
μ
L
Normal and Sequential
1.0–1,000
μ
L
Concentrate and
Sequential Concentrate
100–1,000
μ
L
5 mL
Normal and Sequential
1.0–4,700
μ
L
Concentrate and
Sequential Concentrate
100–4,700
μ
L
Simultaneous
1,000–4,700
μ
L
10 mL
Normal and Sequential
1.0–8,000
μ
L
Concentrate and
Sequential Concentrate
100–8,000
μ
L
Simultaneous
1,000–8,000
μ
L
Table 3-2. Summary of Injection Volume Ranges