2 • Description
Doc. 065051-03 1/08
23
2.3.8
Simultaneous Injection Option
When the AS simultaneous injection option is installed, a single AS can
deliver sample to two independent chromatography systems. The sample
is injected simultaneously to both systems (two injection valves are
required). Dual analyses can be performed with only one sample, thus
increasing sample throughput and eliminating the need to label, fill, and
track two sample vials.
In Chromeleon or Chromeleon Xpress, the two chromatography systems
and the AS are configured into a single timebase and each system is
assigned a unique device name and channel. This lets you monitor and
control both systems from one Control panel and run all samples in one
sequence.
The simultaneous injection option requires a 5 or 10 mL
sample
syringe
and an 8.5 mL sampling needle assembly (see
.
Full-loop
injections are required with this option.
See
for details about simultaneous injections.
2.3.9
Sequential Injection Option
When the simultaneous option is installed, a single AS can deliver sample
to two independent chromatography systems. With the sequential option,
sample is delivered to the first system, flow is rerouted (diverted), and
sample is then delivered to the second system. Sequential injection allows
different samples to be injected into each system.
The sequential injection option requires two injection valves and a
diverter valve (for switching sample flow between the two systems). The
diverter valve is installed in the AS autosampler compartment. The
injection valves must be installed in the IC system’s chromatography
compartment.
See
for details about sequential injections.
NOTE The diverter valve can only be controlled by
Chromeleon or Chromeleon Xpress. It is not
controllable from the AS front panel.
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...