5 • Service
Doc. 065051-03 1/08
165
5.13 Replacing the Sample Syringe Valve
1. Disconnect the liquid lines from the existing valve.
2. Remove the existing valve (see
3. When the valve is removed, the syringe drive shaft is visible (see
). When the syringe is in the
INITIALIZE
position (the position
selected when the valve was removed), the slot in the drive shaft should be in
the position shown in
. The position depends on whether a prep
syringe is installed.
If the slot in the drive shaft is not in the correct position, the syringe drive may
need replacing. Contact Dionex for assistance. Do not attempt to manually
turn the drive shaft.
4. Check the back of the new
sample syringe valve
(P/N 054242). The valve
drive shaft should be
perpendicular to valve ports
A and D (see
If it is not, rotate the valve
rotor until the drive shaft is
in the correct position.
Figure 5-33. Sample Syringe Valve Drive Shaft
Initial Position
Figure 5-32. Sample Syringe Drive Shaft Initial Positions
Sample Syringe
Drive Shaft
Front View of Sample Valve
Drive Shaft
(Initial Position)
Port A (On Front)
Port D (On Front)
Back View
of Valve