3 • Operation and Maintenance
Doc. 065051-03 1/08
97
3.8
Syringe Speed and Tubing Size Guidelines
The AS is capable of delivering at a
maximum pressure of 690 kPa (100 psi)
.
Several factors affect the system backpressure, including the syringe speed,
syringe size, and the IDs and lengths of the sample loop and the tubing used to
connect the inject port and the injection valve. Use the following formula to
determine the backpressure produced by these factors:
backpressure = cm of tubing x kPa (psi)/mL/min/cm
) x mL/min
(from
)
Example:
30 cm of 0.254-mm ID tubing x 1.6 kPa/mL/min/cm x 2.40 mL/min
(5 mL syringe at speed 2) = 115 kPa
Tubing ID
Backpressure in kPa (psi) per
mL/min per cm of tubing
Volume in
μ
L per
cm of tubing
0.127 mm (0.005 in)
25 kPa (3.6 psi)
0.13
0.178 mm (0.007 in)
6.5 kPa (0.94 psi)
0.25
0.254 mm (0.010 in)
1.6 kPa (0.23 psi)
0.51
0.330 mm (0.013 in)
0.55 kPa (0.079 psi)
0.86
0.508 mm (0.020 in)
0.10 kPa (0.014 psi)
2.03
Table 3-10. Tubing Volume and Backpressure
Sample Syringe Size
Sample Syringe Speed
Flow Rate (mL/min)
100
a
, 250, 500, and 1,000
μ
L
a. The maximum flow rate for a 100
μ
L syringe is 0.60 mL/min.
1
0.06
2
0.12
3
0.24
4
0.60
5
1.20
5 and 10 mL
b
b. In general, if you are using a 5 or 10 mL syringe, select a syringe speed of 1 or 2.
1
1.80
2
2.40
3
4.80
4
7.20
5
15.00
Table 3-11. Flow Rates Provided by the Sample Syringe
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...