C • Display Screens
Doc. 065051-03 1/08
245
C.5 Schedule Screen
Use the
SCHEDULE
screen to program a sequence of injections.
Figure C-12. Schedule Screen
Screen Field
Description
Values
EDIT
The schedule number to edit. Press a numeric
button to select the number or press
Select
Δ
or
Select
∇
to scroll through the choices. Press
Enter
to display the selected schedule.
1 to 9
SAVE TO
The schedule number to save the current schedule
to. It does not have to be the
EDIT
number.
1 to 9
RUN
The schedule number to run.
Note
: The schedule
will not start until you press
Hold/Run
or send a
TTL/RLY run command.
1 to 9
MISSING VIAL
ACTION
The action the AS takes if a vial is missing from
the tray. Select
SKIP
to ignore the missing vial and
move to the next vial in the schedule. Select
STOP
to stop the schedule. To resume, press
Hold/Run
.
SKIP/STOP
(default=
STOP
)
LINE
Each line contains the parameters for one vial in
the schedule. Line numbers are assigned
automatically when you insert steps. (A step is
one horizontal line on the
SCHEDULE
screen.) If
consecutive vials have identical parameter values,
line numbers are shown in a range format (see the
example above). A schedule can have a maximum
of 203 steps, containing up to 999 lines.
1 to 999
(automatically
assigned)
Help Prompt
SAVE TO
RUN
SCHEDULE
25.0
5.0
1
1
1
4
3
5
5
1
6
6
1
1
MISSING VIAL ACTION:
INJ/
INJ
VIAL
METHOD
VOL(uL)
VIAL#
LINE
END
START
EDIT
SKIP
1
1
1
1-4
5
6
-
-
-
-
25.0
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...