AS Autosampler Operator’s Manual
68
Doc. 065051-03 1/08
Partial-Loop and Partial-Loop, Limited-Sample Volume Examples
For partial-loop injections, the amount of sample aspirated is equal to the
injection 2
times the cut volume. The amount injected is equal
to the injection volume.
For partial-loop, limited-sample injections, the amount of sample
aspirated and injected is equal to the injection volume.
Maximum Sample Loop Sizes and Injection Volumes
The maximum loop size allowed depends on which sample syringe and
sampling needle assembly are installed. If the sample loop is too large, the
sample volume drawn exceeds the syringe/sampling needle assembly
volume, causing sample to enter the flush bottle.To prevent sample from
contaminating the flush bottle, the sample loop size must not exceed the
values shown in
Installed
Loop Size
(
μ
L)
Injection
Volume from
Chromeleon
Sequence (
μ
L)
Cut
Volume
(
μ
L)
a
a. The cut volume specifies a volume of sample to be “cut” (or omitted) from
each side of the sample during the injection process. See
for
details.
Sample
Aspirated
from Vial or
Well (
μ
L)
Sample
Volume
Injected onto
Column (
μ
L)
25
10
5
20
10
25
20
10
40
20
25
10
0
10
10
25
20
0
20
20
Table 3-8. Partial-Loop and Partial-Loop, Limited-Sample
Sample Volume Examples
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...