C.9.3
XYZ Test Screen
Use the
XYZ TEST
screen to test operation of the sampling needle arm.
INJECT VALVE 1
INJECT VALVE 2
(simultaneous
injection mode)
If the AS is equipped with two injection valves (for
simultaneous injections), these tests check that each valve
operates correctly.
Figure C-28. XYZ Test Screen
Screen Field
Description
ACTION
The action the needle arm should perform:
HOME
Moves all three axes to the home position.
GO TO X
Positions the arm at the specified location
(1, 2, 3, or 4).
CYCLE
Continuously cycles the arm through the four
positions.
PATTERN
Continuously cycles through a four corners test
pattern, with the needle moving about halfway into
the vial.
VIALS
Continuously cycles the arm through each position in
the currently installed tray, with the needle tapping
the top of each vial.
VIALS+I
Continuously cycles the arm through each position in
the currently installed tray, with the needle inserted
into each vial.
ACTUAL XYZ
Displays the needle arm's current XYZ position. The numbers
indicate the number of motor steps away from the home position.
Test
Description
Help Prompt
XYZ TEST
0
0
0
0
0
1
0
0
0
0
0
0
1:
2:
3:
4:
1234
1234
5678
5678
901
901
ACTION:
X SENSOR:
Y SENSOR:
Z SENSOR:
CYCLE
ACTUAL:
X
Y
Z
1
Z HOME:
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...