AS Autosampler Operator’s Manual
50
Doc. 065051-03 1/08
11. Close the autosampler compartment door. Make sure the door
remains closed during operation.
Guidelines for improving precision when sampling from well plates:
•
To avoid contaminants, rinse well plates and covers thoroughly with
deionized water before use.
•
Dry well plates completely to prevent any changes in sample
concentration. You can use 690 kPa (100 psi) nitrogen gas to assist in
drying.
•
Follow the sample volume guidelines
sample into the wells.
•
Check the orientation of the cover to make sure the notches in the
corners of the cover and well plate are aligned.
•
Press the well plate cover tightly into each well.
•
To avoid contaminants and evaporation, do not reuse covers. The
small hole created when the sampling needle pierces the cover allows
slight, but noticeable, evaporation of sample. This is especially true
when samples must wait for several hours before being run.
•
If sample evaporation does occur, try reducing the tray temperature.
If you open the door during operation, the sampling arm stops
immediately. If a schedule is running, it is ended. To safely open the
door during operation, press the Door button (see
for
details).
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...