D • Front Panel Control
Doc. 065051-03 1/08
313
7. Select the following options:
ASPIRATE
:
60
μ
L for normal, simultaneous, and concentrate modes;
120
μ
L
for sequential and sequential concentrate modes (to empty the needle seal line
and diverter tubing)
from
NEEDLE
SYRINGE SPEED
:
3
ACTION
:
ASPIRATE
8. Press
Enter
.
9. Select
VIAL#
:
FLU
and press
Enter
.
The needle arm moves to the flush port.
10. Select
ACTION:
EMPTY
and press
Enter
.
11. Select the following options:
from
FLUSH
SYRINGE SPEED
:
5
ACTION
:
FILL
12. Press
Enter
.
13. Select the following options:
DISPENSE
:
50
μ
L
SYRINGE SPEED
:
5
ACTION
:
DISPENSE
14. Press
Enter
.
15. Select
VIAL#
:
INJ
and press
Enter
.
The needle arm moves to the inject port.
16. Select the following options:
DISPENSE
:
25
μ
L for normal mode;
75
μ
L for sequential mode
SYRINGE SPEED
:
1
ACTION
:
DISPENSE
17. Press
Enter
.
The syringe dispenses 25
μ
L or 75
μ
L.