AS Autosampler Operator’s Manual
252
Doc. 065051-03 1/08
C.6.5
System Parameters Screen
Use the
SYSTEM PARAMETERS
screen to set default system operating
parameters. The parameters vary, depending on whether the AS is
controlled exclusively from the front panel or if it is sometimes or always
under Chromeleon or Chromeleon Xpress control.
Figure C-19. System Parameters Screen:
AS Controlled Exclusively from Front Panel
Figure C-20. System Parameters Screen:
AS Controlled from Chromeleon or Chromeleon Xpress
Screen Field
Description
Values
WAIT
Enables or disables the wait step in the sample prep
phase of the method. When wait is enabled, the AS
waits for a run command from either the front
panel
Hold/Run
button or a TTL input signal
before performing a method’s timed events steps.
This allows another instrument to control the
timing of each injection in a schedule.
ENABLED
DISABLED
(default =
ENABLED
)
Help Prompt
SYSTEM PARAMETERS
WAIT:
ENABLED
METHOD SYRINGE SPEED:
CUT SEGMENT VOLUME:
INJECT PORT VOLUME:
uL
uL
10
S1 43
3
SAMPLE MODE:
NORMAL
FLUSH SYRINGE SPEED:
4
uL
S2
Help Prompt
SYSTEM PARAMETERS
WAIT:
ENABLED
PIPET SYRINGE SPEED:
CUT SEGMENT VOLUME:
INJECT PORT VOLUME:
uL
uL
10
S1 43
3
SAMPLE MODE:
NORMAL
EMPTY SYRINGE SPEED:
4
uL
S2
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...