B.3.13 Priming the Liquid Lines
This procedure explains how to prime the liquid lines from the
Chromeleon or Chromeleon Xpress Control panel. If you prefer, you can
use the
FLUSH
screen on the front panel to prime the lines (see
).
NOTE To reduce bubble formation while priming, degas
the flush solution.
1. Open the Chromeleon or Chromeleon Xpress Control panel (see
2. Enter the
Prime Volume
suggested volumes.
3. If the sample preparation option is installed, select
Prep
in the
Prime
Syringe
list and select
Flush_Reservoir
in the
Prime reservoir
list.
4. Click
Prime
.
The flush reservoir line and sampling needle assembly are primed
(see
). If the sample preparation option is installed, the
reagent holding loop is also primed (see
5. Check for bubbles in the lines and repeat the priming procedure, if
necessary, to eliminate the bubbles.
Figure B-47. Prime Commands on the Control Panel