AS Autosampler Operator’s Manual
298
Doc. 065051-03 1/08
D.5.9
Example Method for Dilution
NOTE The sample preparation option is required for this
method.
This method uses the sample preparation option to dilute a sample before
injection. In this method, the AS delivers 600
μ
L of the concentrated
sample from vial 6 and 5400
μ
L of diluent from reagent reservoir A to the
sample vial. Default values are used for all other method parameters.
shows the sample prep steps for the method.
D.6 Creating Schedules from the AS Front Panel
A schedule defines injection parameters for each vial to be sampled. A schedule
consists of the following parameters:
•
The vials to sample
•
The number of injections to draw from each vial
•
The volume of each injection
•
The method to run on each injection
•
The action to take when a scheduled vial is missing
You can create a new schedule or edit an existing schedule when the
AS
is
running, on hold, or stopped. If the schedule you are editing is currently running,
changes to steps that have not yet been completed are performed in the current
run. Changes made to steps that have already completed are performed the next
time the schedule runs.
Figure D-11. Dilution Method Example: Sample Prep Steps
Help Prompt
METHOD:
SAMPLE PREP
1
2
WAIT
1
600uL
uL
+
# 6
#SMP
DILUTE
5400
A
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...