Troubleshooting
MagnaTran 7.1 User’s Manual
Power Related Issues
MN-003-1600-00
Brooks Automation
10-6
Revision 2.2
Power Related Issues
Symptoms:
24 volt LED not illuminated (located on robot I/O board)
Failure for robot to operate
No communication using personal computer or CDM
Troubleshooting Process:
Check facilities power to power supply.
Verify power supply rating is 24 volt, 30 amperes.
Verify voltage output at power supply is 24 ±2 volts.
Verify voltage at robot power connector is 24 ±2 volts. Refer to
for power connector pin-outs.
Check power cables for proper connection and continuity.
Replace robot I/O board. Refer to
I/O Board Replacement on page 9-48
Call Brooks Technical Support.