Running Samples
Running Samples - MCL Automatic Mode
8
5
or
to select the Worklist, Panel or Protocol
you want to use.
NOTE
If you load a Panel or Worklist that contains a deleted Protocol, the software prompts you to use
Wizard to correct the problem. Panel names for the corrected protocols are not
displayed in the Worklist however they are displayed on the Panels tab.
6
Enter the carousel ID number and tube location in the Acquisition Manager.
7
to begin processing samples.
During the sample cycle the following series of Cytometer status messages appears:
Awaiting Sample
Preparing Sample
Acquiring
Stopping.
8
Observe the Events/Sec counter on the
to monitor data acquisition.
NOTE
While running a Worklist and using bar codes on your sample tubes, if the Tube ID in the Worklist
does not match the bar code on the tube, a system message appears. After you acknowledge the
message, the mismatched tube is aborted and the processing stops for the rest of the carousel.
9
Verify all sample IDs before reporting results. Review all data plots and results, including
exported results, prior to reporting results.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......