773232AH
10-112
Using Navios Software
Plots Menu
In normal blood, where lymphocytes account for about 30% of the white cell count, we have no
problem with the above setup. We can predict with a normal sample that the total count from a
5,000 cell lymphocyte count is around 5,000 x 100/30 or 16,666 events total.
If the sample is from a lymphopenic patient, this could be a problem. A 5,000 lymphocyte count may
not be achieved before the sample tube is empty or RAM memory is full.
It is normal under these circumstances to set two counters, one for the stop count and one for the
maximum permissible count, that is, the maximum total event count. If a stop count is used, it is
recommended that the stop count be set to a large number, for example, 100,000 events. Thus, with
a lymphopenic patient, if the lymphocytes are less than 5% of the white cell count, an acquisition
would terminate on the total event count of 100,000 before a 5,000 cell lymphocyte count could be
achieved.
NOTE
If a Stop condition is reached while you are adjusting voltages and gains, acquisition is not restarted
and the data plots are not refreshed. You can avoid this situation by making all cytosetting adjustments
while in
.
Setting a Stop Counter
NOTE
If a stop count is used on a plot that contains an AutoGate region, the stop count is not exact.
1
H
ighlight the plot on which you wish to set a stop counter.
2
Plots > Acquisition Save and Stops
command to view the Acquisition Plot Set-up dialog
box.
Save Histogram Data in FCS Format
NOTE
You cannot make any changes on this screen when running a
protocol.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......