Replace/Adjust Procedures
Replace An Optical Filter
3
Place the filter (2) in the filter holder and tighten the set screw (1).
Identify Coated Side Of Dichroic Filter
1
Determine which is the coated side
(1 )
of a non-BCI filter:
a.
Take the eraser end of a pencil and hold it close to the filter, near its edge.
b.
Look at the two reflections, dark- and light-colored, of the pencil.
c.
Turn the filter over and repeat steps
.
d.
The side where the pencil touches the dark-colored reflection is the coated side
(1)
.
The uncoated side
(2)
shows the pencil touching the light-colored reflection.
B
C
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......