773232AH
8-10
Running Samples
Running Samples - MCL Manual Mode
11
Observe the Events/Sec counter on the
to monitor data acquisition.
NOTE
While running a Worklist, if the Tube ID in the Worklist does not match the bar code on the tube,
a system message appears. After you acknowledge the message, the mismatched tube is aborted
and the processing stops for the rest of the carousel.
12
If you enabled
Edit
Sample IDs
on
Workspace Preferences - Acquisition Options
, edit the sample
ID after the tube finishes processing and
.
NOTE
After the first tube in a panel, Sample ID 1 cannot be changed.
13
Enter the location of the next tube to process and
.
14
Repeat steps
to process additional tubes.
15
Verify all sample IDs before reporting results. Review all data plots and results, including
exported results prior to reporting results.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......