773232AH
10-168
Using Navios Software
Cytometer Menu
Pause and Rotate
CAUTION
Risk of sample misidentification. Sample misidentification can occur if you use the
Pause and Rotate button and remove the sample tube and replace it with a
different sample tube. To prevent sample misidentification, do not swap sample
tubes when using Pause and Rotate button.
The Pause and Rotate button
allows you to pause processing of a single sample and rotate the
MCL to a fixed position on the carousel so that additional reagent, for example, can be added to the
sample.
NOTE
To prevent excessive air from entering the flowcell, do not leave the Cytometer paused for extended
periods of time.
NOTE
When you use the Pause and Rotate button function, and remain in the Pause and Rotate state, the
Elapsed Time setting limit is ignored.
Idle Mode
The Idle Mode button
places the Cytometer in Idle mode so that various cleaning and replacing
procedures can be performed.
Prime
The Prime button
allows you to declog or remove bubbles or blockages from the sample line
with sheath solution.
Cleanse
The Cleanse
button allows you to initiate a cycle to flush the instrument sample lines with
cleaning solution. The Cytometer will automatically transition to the idle mode
after the 60
second cleanse cycle completes.
Laser Control
The
Laser Control
Dialog allows you to place the Red and Violet lasers into a Standby state to extend
their lifetimes. The Blue laser cannot be placed into Standby mode by the software.
NOTE
If the Laser Control dialog is open when a laser changes state (e.g. from Warming Up to On), the
current state of the laser may not display correctly. Close the Laser Control dialog and re-open it to display
the current status of the lasers.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......