773232AH
4-11
Specifications
Software Specifications
4
keywords that provide additional information about the data being collected. Custom keywords are
preceded by the “@” character.
Below is a list of keywords included in the TEXT part.
Table 4.1
FCS 2.0 Keyword Information
FCS 2.0 Keyword
Key value
$BTIM
Start time (HH:MM:SS 24 hour clock ':' delimiters)
$BYTEORD
Little endian: “1,2'”
$CELLS
Cells name (from Workspace Preferences)
$CYT
Type of flow cytometer: Navios.
$DATATYPE
Single character defining base data type of binary data, I=integer.
$DATE
DD-MMM-YY
$DFCiToj
Compensation matrix. % of FLi to subtract from FLj. Where FLi and FLj refer to
fluorescence detector numbers. The size of the compensation matrix is equal to the
number of fluorescence detectors (colors) of the instrument.
So, for a 10 color instrument there will be 100 values from $DFC1TO1 … $DFC1TO10
and $DFC10TO1 … $DFC10TO10 and all combinations in between. For FL1 through
FL10.
A six color instrument will produce 36 $DFCiTOj values. For FL1 through FL6.
$ETIM
End time (HH:MM:SS 24 hour clock ':' delimiters)
$EXP
Experiment name (from Workspace Preferences)
$FIL
File name excluding path
$INST
Institution name (from Workspace Preferences)
$INSTADDRESS
The address of the facility.
$MODE
Single character defining mode of data, L=listmode.
$NEXTDATA
Byte offset into binary data for next data item. In this case it is the offset to the
beginning of the FCS 3.0 section.
$OP
Operator ID, that is, user ID.
$PAR
Number of parameters selected in the protocol. Including all linear and logarithmic
versions that were selected. This number also indicates the number of parameters
per event in the DATA part. The parameter indices are determined by the order in
which they were selected in the protocol.
$PnB
Number of bits allocated to store data for parameter n: 16 = 10 bit data per
parameter plus 6 bits of padding.
$PnE
Lin/Log parameter n calibration (decades, offset) “0,0” means linear and “ 4.0,
0.1024” means logarithmic. This value is used to specify the offset and decades
used to set the tick mark scaling on plots for log parameters.
$PnG
Amplifier gain used for acquisition of parameter n.
$PnN
Name of parameter n, for example: Manufacturer assigned name, that is, FS INT LIN
or FL3 PEAK LOG
$PnR
Channel resolution of parameter n: 1024 = 10 bit parameters.
$PnS
User assigned stain name of parameter n.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......