Operation Principles
Plot Display
3
Plot Display
The results of sample analysis appear on the Workstation screen as graphs called plots. You assign
the parameters to the plot axes. Plots can be displayed in black and white or color as:
•
Single-parameter histogram plot
•
Dual-parameter dot plot
•
Dual-parameter density plot
•
Prism plot
Dual-parameter plots can be displayed in 64 x 64, 128 x 128, 256 x 256, or 512 x 512 resolution. Plots
can be displayed on a black or white background.
PRISM
Prism is used to analyze multicolor immunofluorescence samples. With multicolor
immunofluorescence a cell is either positive or negative for each of up to ten cell surface markers.
A particular combination is called a phenotype. Prism allows you to display percentages on all
phenotypic populations in a single plot. It is software derived and can be acquired in either run time
or listmode.
Prism is available on up to ten parameters. TIME, RATIO, and Prism itself cannot be used for Prism.
All other signals can be used for Prism. Generally, the fluorescence parameters are used.
A Prism plot shows a spike or population for each antibody combination with a percent of the total
that represents the percent of the total events in the Prism plot. See
.
Figure 3.11
Prism Plot
See
for instructions on using Prism.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......