Troubleshooting
Cytometer Messages Table
Optics Temperature Stabilizing
The temperature of the optics is
stabilizing.
If the optics temperature does not stabilize
within a reasonable time then restart the
entire system. If the problem persists then
contact your local Beckman Coulter
Representative.
Optics Temperature Warning
The temperature of the optics differs
considerably from the temperature
when the optics were aligned such
that the instrument may not meet its
specification.
Wait till the temperature of the optics is
closer to that when the instrument was
aligned or realign the optics for different
ambient temperature range.
Overflow Error
The drip or waste container has
overflowed and the acquistion has
been stopped.
Contact your local Beckman Coulter
Representative.
Pico Motor Board Not Found
The pico motor control board could
not be detected.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Place tube in position 21
Single Tube Fixed Mode selected
Place tube in position 21 and select Run to
continue in this mode.
To discontinue operation in this mode,
depress the Single Tube Fixed Mode button
on the Cytometer Control toolbar.
Pressures Not Calibrated
Pressures can not be calibrated
because pressure control board can
not be detected or calibration
coefficients corrupted.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Probe Up/Down Error
The probe failed to move up or down
correctly after several attempts.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Probe Up/Down Warning
The first attempt to raise or lower the
probe failed.
None. The instrument will automatically
retry to raise or lower the probe. If warning
persists, contact your local Beckman
Coulter Representative.
Red Laser Comms Warning
Unable to communicate with red
laser.
Restart the entire system. If the error
persists, contact your local Beckman
Coulter Representative.
Red Laser Fault
A fault occurred on the red laser
during acquisition and the
acquisition was stopped.
Restart the entire system. If the error
persists, contact your local Beckman
Coulter Representative.
Red Laser Fault Warning
There is a fault condition on the red
laser.
Check optics covers are properly installed
and tightened. See Red Laser Fault.
Red Laser Initialization Error
The red laser failed to initialize.
Restart the entire system. If the error
persists, contact your local Beckman
Coulter Representative.
Table 13.2
Cytometer Messages
(Continued)
Message
Probable Cause
Recommended Action
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......